Adding Systems
Adding an Auxiliary Gas System
540
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
7. Connect the Auxiliary Gas Module electrically.
a.
Using the cable provided, connect the 15-pin female connector marked
GC Bus
on
the module to a
Bus
interface on the back of the GC.
8. Replace the left/right side panel.
a.
Plug the ground wire previously removed into the left/right panel.
b. Place the left/right panel and attach the screw holding it in place.
9. Open the gas supplies.
10. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
11. Power on the GC.
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker), located at the back of the instrument, to the
position I.
IMPORTANT
The maximum nominal inlet pressure for all the inputs is 1000 kPa
(145 psig). The working inlet pressure range is from 400 kPa (58 psig) to 1000 kPa
(145 psig).
IMPORTANT
If two Auxiliary Gas modules are simultaneously present, one must be set as
Primary and the other as Secondary by mean of dedicated switch on the back panel.
See the example below.
The primary module controls the aux pressures from 1 to 3, while the secondary module
controls the aux pressures from 4 to 6.
Secondary
Primary