Performing Detectors Routine Maintenance
Maintaining a Flame Ionization Detector (FID)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
145
6. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
7. Open the module flap cover.
8. Disconnect the signal cable, glow-plug and polarizing cables from their contacts on the
cell top cover.
Figure 120.
FID Cables
Figure 121.
FID Cables Disconnection
a.
Loosen the glow-plug cable socket set screw using a T6 Torxhead screwdriver.
Carefully pull out the terminal body of the glow-plug cable from the dowel pin on
the glow-plug element.
Glow-plug Cable Terminal body
Polarizing Cable Straight Jack Connector
Signal Cable Straight Plug Crimp
Connector
Socket Set Screw
Polarizing Electrode Bulkhead Jack
Collecting Electrode (Anode)
Bulkhead Jack
Glow-plug Dowel Pin