12. Close the module flap cover.
13. If present, move the autosampler towards the module to restore the original alignment.
14. Turn the carrier gas on.
15. Set the normal injector, detector, and GC working conditions.
Replacing a SSLBKF Broken Liner
To replace a broken liner
Figure 86.
SSLBKF Injector: Broken Liner Replacement
CAUTION
Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
Liner
Liner Seal
Retaining Nut
Washer
Base Seal
Ferrule
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
Capillary Column Retaining Nut
Bottom Fittings