9
Adding Modules
Adding a Helium Saver Injector Module
448
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
9. Open the gas supplies.
10. If other external modules are present, plug the power cable to the AC Input connector of
each external module, and to the wall outlet.
11. Power on the GC.
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker), located at the back of the instrument, to the
position I.
12. Configure and enable the system installed through the user interface of your GC, or
through the CDS in use. Refer to the
TRACE 1300 and TRACE 1310 User Guide
.
Adding a Helium Saver Injector Module
The Instant Connect Helium Saver Injector Module (HeS-S/SL) is designed as a “double
wide” module that fits into the space provided by one detector module and one injector
module on the upper deck of your TRACE 1300/TRACE 1310 GC.
Installing the Module
The module may be placed in either the Front or Back position on the upper deck.
• If the module is located in the
Front
position, plumb the helium supply to the Front
Make Up gas connection on the back panel of the GC.
• If the module is located in the
Back
position, plumb the helium to the Back Make Up gas
connection on the back panel of the GC.
Plumb the nitrogen gas to the corresponding Front or Back Carrier gas connections.
IMPORTANT
The Auxiliary Temperature/Cryo module must be powered at the same
power supply voltage required for the GC.
IMPORTANT
Although nitrogen will not be the actual carrier gas for the analytical
separation, it will be the carrier gas during the injection, and it is necessary to plumb the
nitrogen to the carrier gas input.