Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
178
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Removing the FPD Detector
To remove the FPD detector assembly from the FPD detector base body
1. Put the GC in standby condition.
2. Switch off the flame. Hydrogen and air are automatically closed.
3. Cool the detector to room temperature.
4. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If other external modules are present, unplug the power cable from the AC Input
connector of each module, and from the wall outlet.
6. Remove the FPD detector from the base.
a.
Disconnect the signal, excitation voltage, and ignition/heating cables from the
detector. See
Figure 160.
FPD Cable Disconnection
b. Using the tool provided with the system, loosen the fixing nut on the base of the
.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Signal Cable
Excitation
Voltage
Cable
Ignition/Heating Cable