Injectors Advanced Maintenance
Cleaning the SSLBKF Injector Body
238
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Cleaning the SSLBKF Injector Body
Perform this operation when a more efficient cleaning of the injector body is desired or
required, due to contaminants.
To clean the injector body
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
4. Put the autosampler away if present.
5. Open the module flap cover.
6. Remove the top parts of the injector. See
.
a.
Unscrew the septum cap of the injector.
b. Unscrew the ring nut.
c.
Remove the septum holder/liner cap with septum from the injector body head.
7. Remove the liner. See
a.
Use tweezers to remove the liner with the liner seal from the injector.
8. Remove the bottom parts of the injector. See
.
a.
Unscrew the capillary column retaining nut, then remove the analytical column with
its ferrule from the bottom of the injector.
b. Unscrew the retaining nut with the washer and the base seal.
Materials needed
Ultrasonic cleaning bath
Methanol/acetone mixture 1:1
GC-grade methanol
1/8-inch -wrench
T20 Torxhead screwdriver
Forceps or tweezers
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.