Connecting the Carry Bar to the Lift Strap

1.

Insert the G

-Series 

Lift strap loop into the the top of the hanger assembly, 

making sure that the pin hole lines up with the loop.

2.

Next Insert hanger assembly pin through the making sure is it passes through 

the loop and to the far side of hanger assembly.

3.

Finally, attach the safety ring on the end of the pin making sure it won't come 

free.

!

Do not pull the red cord forcefully. For assistance after an emergency, contact 

Vancare

 

Service

 

at 1.800.694.4525

Manual Emergency Lowering

The manual emergency and raising should only be used if the emergency lowering does not work. A proper safety ladder or 

stool may be required to remove the plug from the cover.

Remove the round plug from the lift cover and use 3/16" or 2.5 mm Allen key to rotate the motor in the up or down direction.

!

Caution: DO NOT attempt to use the lift while using manual lowering.

Disconnecting the G-Series Lift from the Trolley

To disconnect the lift from the Trolley, lift the G

-Series 

Lift then press on the two Quick Release buttons and pull 

down on the lift.

Quick Reference Guide

Operating the G-Series Lift

The

 

Vancare

 

G

-Series 

Lift is switched ON automatically when a button on the hand control 

is pressed. It will switch OFF automatically after approximately three minutes without 

activation.  The Magnetic Hand Control should be placed on the Carry Bar when not in use.

Basics in Transferring a Patient

The Lifting Sling:

 

A lifting sling with four to six straps designed for mounting on hooks should be used when using an Vancare 

G

-Series 

Lift Carry Bar. Place the straps on the hooks and make sure the loops are not in the latch so the straps do not 

unintentionally fall off.

Working with the 

Vanc

a

re

 G

-Series 

Lift:

 

The Vancare G

-Series 

Lift moves freely in the track system and does not have any 

special requirements for space or power in connection with moving. Attention can be fully focused on the user’s functional level 

and the caregivers  technique. The patient should only be lifted to the extent that she/he is clear of the surface and should be 

moved at this height.

Attaching the Lifting Sling:

 

Place the straps from the lifting sling on the hooks of the hanger assembly. Start with the 

uppermost set of straps (from the back) and then take the lowest set of straps (from the legs).

Emergency Stop

The G

-Series 

Lift unit also has an Emergency Shut-off feature that allows the operator to shut the power to the lift completely in 

the event of an emergency. By pulling once on the RED emergency lowering cord, located on the underside of the lift, the lift 

will immediately stop and all its functions will be disabled.  The “ON” indicator light will turn off, and the Emergency Shut-off 

button located inside the lift will pop out. After an emergency, the lift must be inspected prior to restoring to use. In order to 

restore power to the lift, the tab must be pressed back into the lift.

NOTE: 

Please refer to the complete Installation and Operation Instructions Manual 

for maintenance and safe operation of the equipment.

 www.

vancare

.com  

   

1.800.694.4525

  

"UP" button

"DOWN" button

Magnetic Hand Control

G-Series