6
Injectors Advanced Maintenance
Cleaning the SSL Injector Body
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
237
13. Reinstall the bottom parts of the injector.
a.
Reinstall the retaining nut with the washer and the base seal.
b. Reinstall the analytical column.
14. Reinstall the liner
a.
Using tweezers, place the liner seal over the liner, insert the liner into the injector, and
push it gently towards the bottom of the injector.
15. Reinstall the top parts of the injector.
a.
Place the septum holder/liner cap with the septum on the body head of the injector.
Fix them by screwing the ring nut.
b. Screw and tighten the injector cap to finger tight.
16. Close the module flap cover.
17. If present, move the autosampler towards the module to restore the original alignment.
18. Turn the carrier gas on.
19. Set the normal oven and injector working conditions.
CAUTION
Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.