Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)
112
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
8. Close the module flap cover.
9. If present, move the autosampler towards the module to restore the original alignment.
10. Wait two or three minutes for helium to purge the inlet.
11. Turn the inlet flow back on and enable the inlet heater.
12. Set the normal injector, detector, and GC working conditions.
Cleaning or Replacing the Glass Liner
To clean or replace the glass liner
Figure 99.
HeS-S/SL Injector: Glass Liner Replacement
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
CAUTION
Do not overtighten the septum cap because you might damage the septum and
affect instrument performance.
Note
Care should be taken when performing the procedure in order to keep from
damaging analytical columns.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Septum Cap
Ring Nut
Septum Holder/Liner Cap with Septum
Liner Seal (O-ring)
Liner