9
Adding Modules
Adding a PDD Module
470
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Installing the Capillary Column
For PDD, the column must penetrate
136 mm
inside the pre-installed capillary column
adapter.
To install the capillary column
1. Make a mark on the column 136 mm from the end.
2. Remove the knurled nut column inlet at the bottom of the detector. Slide the nut overt
the end of the column, followed by the appropriate column ferrule.
3. Seat the ferrule in the detail of the column adapter and begin sliding the column through
the capillary column adapter and into the column inlet.
4. Get the nut started on the threads and tighten it until you feel it contact the ferrule, then
back off half a turn.
5. Slide the column into the column inlet until the mark is flush with the surface of the
knurled nut, and secure the column in the adapter by tightening the knurled nut finger
tight only.
Installing the Packed Column
To install the packed column
1. Remove the pre-installed capillary column adapter.
2. Replace pre-installed capillary column adapter with the
packed columns adapter
that
penetrates into the PDD cell for the correct length.
3. Connect the packed column to the packed column adapter.
Testing for Leaks
It is critical for the system to be leak-tight. Leak test is strongly recommended before
operating with PDD.
To test for leaks
1. Open the discharge gas supply.
2. Cap the tube and pressurize the entire system with helium to 138 kPa (20 psi).
3. If the system does not hold pressure, check all the fittings with an electronic helium leak
detector. DO NOT use leak detecting liquids.
4. Tighten fittings as required.