Detectors Advanced Maintenance
Measuring the FID Gas Flows
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
267
Figure 245.
FID Top Parts Removal
a.
Unscrew and remove the detector cap, paying attention to not rotate the cell top
cover.
b. Remove the cell top cover and put it in a safe place. Do not damage the FID
collecting electrode pin.
10. Remove the electrode. See
a.
Using forceps or tweezers, carefully extract the collecting electrode, the insulator ring
and the collector insulator through the top of the detector cell. Place all the parts
removed on a clean surface.
Detector Cap
Cell Top Cover
Collecting Electrode
Insulator Ring
Collector Insulator