5
GC Main Frame Advanced Maintenance
Replacing the Flap Motor
228
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
13. Reconnect the power plug to connector
J10
on the backplane board.
14. Reinstall the back cover.
a.
Reconnect the ground wire to the back cover terminal.
b. Replace the cover proceeding in the reverse order in which it was removed.
15. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
16. Power on the GC.
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker) to the position I.
17. Set the normal injector, detector and GC working conditions.