9
Adding Modules
Adding a NPD Detector Module
438
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
9. Install the NPD Thermionic Source Power Module.
The module should be installed into a
n
external module housing provided on the back
of the GC. See
.
Figure 452.
External Modules Housing
10. Remove the cover of the external modules housing where installing the module. See
Figure 453.
Housing Cover Removal
a.
Using a T20 Torxhead screwdriver, unscrew and remove the left or right housing
cover screws.
b. Remove the covers from the housing.
11. Install the module into the housing.
a.
Loosen the two hexagonal screws under the module. See
.
External Modules Housing
Left/Right Housing Covers
Left Cover
Screws
Right Cover
Screws