9
Adding Modules
Adding a GSV Injector Module
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
427
a.
Open the module flap cover.
b. Keeping the module flap cover open, place the module in its seat. Be sure to insert the
25-pin male connector, on the bottom of the module, into the 25-pin female
connector on the detector seat of the detector housing.
c.
Use a T20 screwdriver to tighten the three captive fixing screws without
overtightening.
d. Close the module flap cover.
10. Connect the analytical column end insert it as far as goes and withdrawn about 2 -3 mm.
See the section
“Installing the Column the First Time”
.
11. Close the front door of the GC.
12. Remove the protective caps from the Sample In and Sample Out fittings. See
Figure 442.
Protective Caps
13. Connect the Sample In and Sample Out lines.
a.
By using the proper 1/8-in. tubing, nut and ferrule, connect the inlet sample line to
the
Sample IN
port on the GSV module. See
.
Note
Tighten the center screw first and then secure the side screws.
Note
If you need the
backflush
, connect the
backflush system
into the GC oven
following the instructions provided in the section
Connecting the
GSV
Backflush System
in Chapter 8 of the
TRACE 1300/TRACE 1310 Hardware
Manual
.