6
Injectors Advanced Maintenance
Cleaning the PTVBKF Injector Head Assembly
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
253
Cleaning the PTVBKF Injector Head Assembly
Perform this operation when cleaning the injector head assembly is desired or required, due to
contaminants.
To clean the injector head assembly
1. Put the GC in standby condition.
2. Cool the oven, injectors, and detectors to room temperature.
3. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
4. If external modules are present, unplug the power cable from the AC Input connector of
each external module, and from the wall outlet.
5. Close the gas supplies.
6. Put the autosampler away if present.
7. Open the module flap cover.
8. Remove the top parts of the injector. See
.
a.
Unscrew the septum cap of the injector.
b. Using tweezers, remove the septum from the top of the injector head assembly.
9. Remove the liner. See
WARNING
This operation must be carried out by authorized and trained Thermo Fisher
Scientific Field Service Engineers.
Materials needed
Ultrasonic cleaning bath
GC-grade Methanol/acetone mixture 1:1
1/8-inch -wrench
5.5 x 25 mm Slotted Stubby Driver
T10 Torxhead screwdriver
Forceps or tweezers
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.