Installing Optional Kits
Connecting a SSL/PTV Backflush System for High Temperature
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
369
To connect backflush line, precolumn, and capillary column
The result of this operation is schematically shown in
Note
If the Tee connector should not be assembled into its clamp, proceed as follows:
1. Completely screw the nut on the threaded shaft of the Tee connector.
−
Note that the clamp has two holes: the smaller one is threaded while the bigger other
is not threaded.
−
2. Pay attention to insert the threaded shaft of the Tee connector into the clamp through
the bigger hole not threaded up to grip the threaded hole.
3. Turn the Tee connector until the threaded shaft begins to protrude the clamp.
−
4. Move the Tee connector support on the lower part of the mounting bracket, then
finger-tighten the nut up to grip the clamp without overtightening. This allow to
adjust the position of the Tee connector during the installation of backflush line,
precolumn, and capillary column.
Retaining Nut
Clamp
Tee Connector
Threaded Shaft
Threaded Hole
Not Threaded Hole