Performing Injectors Routine Maintenance
Maintaining a Split/Splitless Injector with Backflush (SSLBKF)
100
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
4. Put the autosampler away if present.
5. Open the module flap cover.
6. Replace the filter.
a.
Remove the filter to replace from its seat by turning it counter-clockwise.
b. Install the new filter, with o-rings, in its seat by turning it clockwise.
7. Close the module flap cover.
8. If present, move the autosampler towards the module to restore the original alignment.
9. Turn the carrier gas on.
10. Set the normal injector, detector, and GC working conditions.
Replacing the SSLBKF Body Head O-Rings
To replace the body head o-rings
Figure 88.
SSLBKF Injector: Body Head O-Rings Replacement
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Body Head Internal O-ring (Carrier gas Line)
Body Head External O-ring (Purge gas Line)