2
Performing Routine Maintenance
Powering On the TRACE 1300/TRACE 1310
70
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
5. If external modules are present, plug the power cable to the AC Input connector of each
module, and to the wall outlet.
6. Flip up the power switch, located on the back side of the GC, to the On (up) position
marked
I
.
a.
TRACE 1300
— When the GC powers on, all the LEDs on the status panel light up
simultaneously, afterward the
Power
light becomes a solid green while all the other
lights turn off. The GC is now in stand-by status. See
Figure 71.
TRACE 1300 Status Panel at the GC Power On
.
b.
TRACE 1310
— Check the main menu appears on the touch screen. See
.
Figure 72.
TRACE 1310 Touch Screen
7. Open the Thermo Scientific Chromatography Data System installed on the computer.
8. Set the analytical parameters.
Power On Status
Stand-by Status