10
Adding Systems
Adding an Auxiliary Gas System
530
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
a.
Remove the partial cut shaped plate from the exterior wall of the oven box for
accessing the insulating material. See
.
Figure 576.
Perform the Duct for the Auxiliary Gas Interface (1)
b. Using a knife or similar tool, gently cut the insulating material following the border.
See
Figure 577.
Perform the Duct for the Auxiliary Gas Interface (2)
c.
Save the removed insulating material in a safe place because it can be reused.
d. On the left side wall in the interior of the oven, remove the partial cut plate from the
corresponding duct. See
Partial Cut Shaped Plate