Detectors Advanced Maintenance
Measuring the NPD Gas Flows
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
279
b. Using the T10 Torxhead screwdriver, remove the three T10 Torx screws from the
thermionic source assembly. See
Figure 258.
NPD Thermionic Source Removal (2)
c.
Gently lift up the thermionic source connector guiding the flexible cable from the
thermionic source support, then remove the thermionic source assemble from the
NPD body. Avoid bumping the bead on the sides of the collector. See
.
Figure 259.
NPD Thermionic Source Removal (3)
T10 Torx Screw
Thermionic Source Connector
Thermionic Source Support
Thermionic Source Assemble
Thermionic Source Flexible Cable
NPD Body
Collecting Electrode Pin