Performing Detectors Routine Maintenance
Maintaining a Nitrogen Phosphorous Detector (NPD)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
157
9. Remove the thermionic source assembly cable.
a.
Twist the ring to disconnect the thermionic source assembly cable.
b. Push and twist the lock so that the button slides up in the groove, then pull the cable
ends apart. See
Figure 137.
NPD Thermionic Source Cable Removal
10. Unscrew and pull out the straight jack connector of the collecting electrode cable from
the collecting electrode bulkhead jack. See
.
Figure 138.
Collecting Electrode Cable Removal
Thermionic Source Assembly Cable
Thermionic Source Assembly
Collecting Electrode Cable
Collecting Electrode Straight Jack
Connector