7
Detectors Advanced Maintenance
Measuring the NPD Gas Flows
278
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 256.
NPD Collecting Electrode Cable Removal
11. Remove the thermionic source
a.
Using a T6 Torxhead key, loosen the dowel which fix the thermionic source
connector to the thermionic source support. See
.
Figure 257.
NPD Thermionic Source Removal (1)
Collecting Electrode Cable
Collecting Electrode Straight Jack
Connector
CAUTION
The thermionic source is delicate. Be careful not to break or crack the source.
When performing maintenance on the NPD, avoid touching the source with your fingers,
and prevent it from coming in contact with other surfaces.
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
Thermionic Source Connector
Dowel
Thermionic Source Support