1
Installation
Making the Gas Supply Plumbing Connections
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
11
Figure 12.
FPD Detector Gas Lines Connection
f.
Be sure to complete all inlet and detector connections before turning on the gas
supply.
3. Supply the gas lines.
Back Carrier Gas Inlet
IMPORTANT
The maximum nominal inlet pressure for all the inputs is 1000 kPa
(145 psig), as indicated on the label under the gas inlets ports on the back of the GC.
The working inlet pressure range is from 400 kPa (58 psig) to 1000 kPa (145 psig).
Note
If auxiliary carrier gases are required, the installation of the
Auxiliary Gas
module is
required. For the installation details, see the section