Adding Modules
Adding a Helium Saver Injector Module
450
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 467.
Removing the Blanking Washers
4. Ensure there is an o-ring present in the o-ring seat of each position on the pneumatic
network block. See
Figure 468.
Confirming Presence of O-rings
5. Plumb a cylinder of high purity helium 99.999% to the appropriate make-up gas
connection on the back of the GC using the adapter provided. See
The Thermo Scientific gas purification kit part number 1R120577-0001 contains a
heated gettering trap which can be used directly with this fitting.
Blanking
Washer
Make-up Gas Port
Delivering Helium