Detectors Advanced Maintenance
Measuring the NPD Gas Flows
282
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
a.
Using forceps or tweezers, reinstall the insulator into the detector body.
16. Reinstall the insulator and the collecting electrode.
a.
Using forceps or tweezers, reinstall the insulator into the detector body.
b. Place the collecting electrode in its housing.
c.
Using a 6-mm wrench, screw the collecting electrode pin into its seat on the detector
body.
d. Screw the straight plug crimp connector of the collecting electrode cable to the
collecting electrode bulkhead jack.
17. Reinstall the thermionic source.
a.
Remove the protective cap covering the new thermionic source.
b. Mount and guide the new source assembly on the NPD body proceeding in the
reverse order in which it was removed. Be careful not to bump the bead on the sides
of the body and collecting electrode.
c.
Rotate and align the thermionic source connector, then tighten the dowel to fix the
connector to the support using the T6 Torxhead key. See
.
Figure 263.
Reinstall Thermionic Source (1)
18. Reconnect and screw the straight jack connector of the collecting electrode cable to the
collecting electrode bulkhead jack. See
Dowel
Correct Alignment