9
Adding Modules
Adding a SSL, SSLBKF, PTV, or PTVBKF Injector Module
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
423
6. Put the autosampler away if present.
7. Remove the dummy module from the position where the injector module will be
installed.
a.
Open the module flap cover.
b. Using a T20 Torxhead screwdriver, unscrew the three captive fixing screws.
c.
Keeping the dummy module flap cover open, lift up the module from its seat into the
injector housing. Place the dummy module on a clean surface.
d. Remove the gas block plug from the gas connections by unscrewing its fixing screw
using a T20 Torxhead screwdriver. See
Figure 439.
Injector Gas Block Plug
8. Open the front door of the GC.
9. Plug the injector module into the main frame.
a.
Open the module flap cover.
b. Keeping the module flap cover open, place the module in its seat. Be sure to insert the
25-pin male connector, on the bottom of the module, into the 25-pin female
connector on the injector seat of the injector housing.
Gas Connection Without Gas Block Plug
Gas Block Plug on Injector Seat
O-ring
WARNING
Make sure the o-ring is placed into its seat on the gas connection.
See
Do not install the module if the o-ring is missing.