1
Installation
Coupling to a Mass Spectrometer
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
25
c.
On the interior of the oven box remove the partial cut plate from the corresponding
duct. See
Figure 30.
Ducts for the Coupling with ISQ Series, TSQ 8000 Series, DSQ II and ITQ Mass
Spectrometer
d. Place the left panel for your MS and attach the screw holding it in place.
3. Introduce the transfer line inner tube into the oven through the duct provided.
4. Attach the transfer line to the GC column using the proper nut and ferrule.
5. Close the front door of the GC.
6. For DSQ II and ITQ, connect the transfer line heater to the
Aux Temperature/Cryo
Module
. See
.
IMPORTANT
Save the pre-shaped plug of insulating material in a safe place because it
could be reused.
ISQ Series and TSQ 8000
Series
DSQ II
ITQ
CAUTION - INSTRUMENT DAMAGE:
Make sure that the GC column has been
conditioned before connecting it to the transfer line. The material released from the
column, (column bleed), during conditioning may contaminate the detector.
CAUTION
An extension cable transfer line is needed when connecting an ITQ mass
spectrometer to the Aux Temperature/Cryo module.