1
Installation
Coupling to a Mass Spectrometer
26
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 31.
Transfer Line MS Heaters
7. To tune and set the ISQ Series, TSQ 8000 Series, DSQ II, and ITQ working conditions,
refer to the
relevant
User Guide
and
Hardware Manual
.
Making the Duct to Couple with a TSQ Quantum Mass Spectrometer
To make the duct for the transfer line
1. Remove the right side panel.
a.
Open the front door of the GC.
b. Use a T20 Torxhead screwdriver to loosen the right side panel screw from the interior
front panel. See
. Save the screw because it will be reused later.
Figure 32.
Right Side Panel Fixing Screw
c.
Slide the panel towards the back of the instrument up to the stop.
Heater Connection for
Transfer Line 1
Heater Connection for
Transfer Line 2