3
Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)
114
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
16. Turn the inlet flow back on and enable the inlet heater.
17. Set the normal injector, detector, and GC working conditions.
Replacing the Carrier and Split Lines Filters
To replace the active carbon filters on carrier gas line and split line
Figure 100.
HeS-S/SL Injector: Active Filters Replacement
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Ensure the
Enable conservation
field in the configuration page of the inlet reads
Yes
.
4. Check that there is at least 20 mL/min split flow exiting the split line, then turn the
column flow to
Off
in the GC user interface and allow the inlet to depressurize.
CAUTION
The dimensions of the filters are different. The filter on the split gas line is
bigger than the filter on the carrier gas line. Do not invert their position when you replace
them. It is not necessary to replace the filters together.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Split Gas Line Active Carbon Filter
Carrier Gas Line Active Carbon Filter