11
Upgrade Equipment
Upgrading a TRACE 1300 to a TRACE 1310
550
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Upgrading a TRACE 1300 to a TRACE 1310
This section provides the instruction to upgrade a TRACE 1300 to a TRACE 1310 GC.
.
To upgrade a TRACE 1300 to a TRACE 1310 GC
1. Put the GC in standby condition.
2. Cool the oven, injectors and detectors to room temperature.
3. Close the gas supplies.
4. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If external modules are present, plug the power cable to the AC Input connector of each
external module, and to the wall outlet.
6. Remove the door cover of the TRACE 1300.
a.
Open the front door and look for the fixing screws that secure the cover and the
handle to the door. See
.
WARNING
This operation must be carried out by authorized and trained Thermo Fisher
Scientific Service Field Engineers.