Performing Detectors Routine Maintenance
Maintaining a Nitrogen Phosphorous Detector (NPD)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
153
b. Using the T10 Torxhead screwdriver, remove the three T10 Torx screws from the
thermionic source assembly. See
Figure 130.
NPD Thermionic Source Removal (2)
c.
Gently lift up the thermionic source connector guiding the flexible cable from the
thermionic source support, then remove the thermionic source assembly from the
NPD body. Avoid bumping the bead on the sides of the collector. See
.
Figure 131.
NPD Thermionic Source Removal (3)
T10 Torx Screw
Thermionic Source Connector
Thermionic Source Support
Thermionic Source Assembly
Thermionic Source Flexible Cable
NPD Body