Installing Optional Kits
Installing the Packed Column Adapters
340
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 340.
Adapter Installation on SSL Injector Bottom
a.
Finger-tighten the retaining nut of the adapter until it start to grip the bottom of the
injector.
b. Use the 1/2 wrench to firmly tighten the retaining nut of the adapter with the base
seal from the bottom of the injector.
7. Remove the bottom parts of the FID, NPD, ECD, TCD, or FPD detector.
See
Figure 341.
Detector Bottom Parts Removal
a.
Using the 1/2 wrench provided, unscrew the retaining nut from the bottom of the
detector.
Note
Save the bottom parts of the detector in a safe place because will be reused
when you restore the original configuration for capillary columns.
Retaining Nut
Adapter for 1/16-in. OD Packed Column
Adapter for 1/8-in. OD Packed Column
Washer
Detector Bottom
Retaining Nut