8
Installing Optional Kits
Installing the NoVent Microfluidics
388
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 393.
Pulling the Column Back from the Transfer Line Nut
15. Loosen the transfer line nut from the column measuring tool.
16. Remove the column, transfer line nut and ferrule from the column measuring tool,
making sure not to move the septum from its location on the column.
17. Insert the column into the transfer line.
Figure 394.
Inserting the Column into the Transfer Line
18. Tighten the transfer line nut until it is just secure enough so that you cannot move it.
19. Loosen the nut by turning it exactly 1 flat backward.
20. Position the column in the transfer line. Use the septum as a guide to measure the correct
length you should insert the column. Be careful not to change the location of the septum
on the column.
Transfer Line
Nut
Septum
Column
Measuring Tool
Column
Transfer Line
Column
Transfer Line
Nut
Ferrule
Septum