Installing Optional Kits
Performing the Dual FPD Detector Configuration
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
409
Performing the Dual FPD Detector Configuration
This section provides the instructions for performing the Dual FPD configuration by
connecting a second photomultiplier tube with a different interferential filter on the same
detector control module. See
.
Figure 420.
Dual FPD Configuration
To perform the Dual FPD configuration you need the following devices:
• Two Flame Photometric Detector modules
• A Dual FPD upgrade kit PN 19050783
The operation consists of the following steps:
• Remotion of the
front
and
back
FPD Detectors from the relevant detector control
module.
• Disassembling the photomultiplier tube from a FPD detector.
• Assembling the second photomultiplier tube on the other FPD detector to obtain the
Dual FPD configuration.
• Remotion of the
jet
from the
front
detector control module.
• Installation of the
PT100
simulator
, provided with the kit, on the detector base of the
front detector control module.
IMPORTANT
The Dual FPD must be installed onto the FPD control module placed into
the
back
site of the detectors housing.
The FPD control module placed into the
front
site of the detector housing must be
equipped with the
PT100 simulator
, provided with the kit, instead of the FPD detector,
and will supply the second photomultiplier of the configuration.