Installing Optional Kits
Installing the Packed Column Adapters
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
339
To install the packed column adapters
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Turn the carrier gas off, and wait for the carrier pressure to go to zero.
4. Remove the analytical column.
a.
Open the front door of the GC.
b. Loosen the retaining nut from the injector and detector fitting on the upper interior
wall of the GC oven.
c.
Remove the analytical column with its nut and ferrule from the bottom of the
injector and the detector.
5. Remove the bottom parts of the SSL injector. See
Figure 339.
SSL Injector Bottom Parts Removal
a.
Using the 1/2 wrench provided, unscrew the retaining nut with the base seal and the
washer from the bottom of the injector.
6. Install the adapter on the bottom the injector interposing the washer. See
.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
CAUTION
Make sure that the liner does not come out from the bottom of the
injector.
Save the bottom parts of the SSL injector in a safe place because will be reused
when you restore the original configuration for capillary columns.
Base Seal
Washer
Retaining Nut
SSL Injector Bottom