Chapter 9
Page no. 1236
JC-DR-A-224.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A224 - Gas Spring for Rotation Worm Gear
6
PROCEDURE
1. Release the four allen screws (2) securing the mounting plate (3) of the Gas Spring (4 mm allen
wrench).
2. Carefully remove the mounting plate (3) making sure that the felt disk (4) remains in place.
CAUTION
Take care to ensure that the old Gas Spring does not fall out of the mounting plate, during
the handling of the mounting plate.
2
2
3
4
4