GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 905
Chapter 9
ERR-SUB-A-003.fm
142/011
INHIBIT
Cannot reach
requested filter posi-
tion
Filter positioning error
During filter position change from Mo to Rh or vice
versa, the control board checks the transition of opti-
cal sensor. If the optical sensor signal is not received
at expected position, this error is produced.
Possible reasons:
1. Filter gear loose
2. Optical sensor failure
3. Control board failure
Actions to correct problem:
1. Change Collimator Set. Refer to
2. Perform Collimator Format Calibration.
D/R A280
CAL A029
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card