GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card DIAG A1000 - Gantry Diagnostic
Page no. 955
Chapter 9
JC-DIAG-A-1000.fm
2-23 JC DIAG A1117 ROTATION COMMUNICATION FAILURE
Diagnostic for:
112/039 error code (CAN code is 0x8210) invalid PDO length
•
Change the Rotation Board (
Job Card D/R A221 - Rotation Board
•
If the error persists, change the PL 103 CPU Board (see
Job Card D/R A190 - Gantry CPU Board
•
If changing the CPU Board solved the error, put back the original Rotation Board (
).
FRUs:
2-24 JC DIAG A1118 ROTATION CONTROL FAILURE
Diagnostic for:
112/045 error code (CAN code is 0x5028) optical fork reached error
•
Inspect the optical fork cable. If the optical fork cable seems to be good, change the optical fork (see
Job Card D/R A223 - Rotation Optical Fork
•
If the error persists, put back the original optical fork and change the Rotation Board (
).
•
Reboot the Gantry. If the error is still present, put back the original Rotation Board as it is good (
Card D/R A221 - Rotation Board
). Then change the CPU Board (see
FRUs:
Rotation Board
Nephtys Gantry CPU
Rotation Board
Nephtys Gantry CPU
Optical fork + cable