Chapter 9
Page no. 1718
JC-CAL-A-049.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A049 - Compression Paddle Offset Calibration
A list of paddle offets similar to the following appears.
---------------------------------------------------
No - Paddle Deformation Offset (1/10th mm))
---------------------------------------------------
1 - Phlex 19x23 0
2 - Std 19x23 0
3 - Auxiliary 0
4 - Round spot 0
5 - Square spot 0
6 - Biopsy 0
7 - Spot biopsy 0
11 - High wall 0
12 - Std mag stand 0
14 - Round mag stand 0
15 - Square mag stand 0
16 - Phlex 19x23 sliding 0
17 - Std 19x23 sliding 0
18 - Round spot sliding 0
19 - Square spot sliding -3
20 - Biopsy sliding 0
21 - Spot biopsy sliding 0
22 - Std 24x30 0
23 - Phlex 24x30 0
4. If any of the final offset values are not as expected, repeat procedure from step
of Section
Determine the Current Offset for Each of the Compression Paddles
for the appropriate
Compression Paddle(s).
5. For each Compression Paddle you changed the offset for, follow these steps:
a. Insert the compression Paddle.
b. Apply a compresison of 10 dAN.
c. Verify the Gantry LCD Readout displays exactly the known thickness of the 40 mm acrylic plate,
as measured in step
7-2 Determine the Current Offset for Each of the Compression
.
If the LCD readout does not display exactly the known thickness of the 40 mm acrylic plate,
repeat the calibration procedure starting from step
7-2 Determine the Current Offset
for Each of the Compression Paddles
6. If all of the offset values are as expected, exit the telnet connection with the Gantry CPU Board, by
entering:
logout