Chapter 7
Page no. 740
JC-ELE-A-043.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
Brightness/Contrast
is selected, and press
the
UP
button
to enter the
RGB
menu.
The
RGB menu
appears.
4. Perform a Monitor brightness adjustment, as follows:
a. Stay in the
RGB
menu, and keep pressing the
MENU
button until
Brightness
is selected.
b. With the luminance meter positioned on the central black level square of the SMPTE pattern (see
the image in
), use the
DOWN
button to decrease the
Brightness
value until the lumi-
nance measured stops changing. Then, use the
UP
/
DOWN
buttons to verify that the luminance
measured is at the limit between not changing and decreasing (see the image below which illus-
trates this adjustment).
5. Perform a Monitor contrast adjustment, as follows:
a. Stay in the
RGB
menu, and keep pressing the
MENU
button until
Contrast
is selected.
b. With the luminance meter positioned on the central white level square of the SMPTE pattern (see
the image in
), use the
UP
button to adjust the
Contrast
value until the luminance mea-
sured stops changing. Then use the
UP/DOWN
buttons to verify that the luminance measured is
at the limit between not changing and increasing (see the image below which illustrates this
Measured lumniance
(on black level square)
Decreasing brightness
(Brightness value displayed on Monitor OSD)
Down
Up/Down
buttons to find limit where
measured liminance is at the limit be-
tween not changing and decreasing.