Chapter 9
Page no. 1260
JC-DR-A-232.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A232 - Lift Screw Assembly
tighten the belt a little to check the correctness of nylon belt installation.
If you do have a Gantry LOTO tool, use the following sub-steps:
a. If necessary (i.e. if the Lift upper limit is not at the maximum setting), manually raise the lift posi-
tion by rotating the Lift Screw anti-clockwise (13 mm or 17 mm hexagonal wrench) to the correct
height, so that the disk of the Arm can be accommodated by the recess of located on the Gantry
LOTO Tool.
One belt end
starts here
Route behind
X-Plate
Other belt end
finishes here
Loop belt here
Bottom part of the
recess
.
Bottom part of the
disk
must be slightly higher
than the bottom part of
the
recess
on the rear of
the Gantry LOTO Tool.