Chapter 9
Page no. 1584
JC-DR-A-408.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A408 - Generator Supply Command Board 200-PL2
S2 before proceeding.
WARNING
Residual voltage stored in the resonant capacitors can be dangerous. Ensure that the res-
onant capacitors are fully discharged before you handle the Generator Supply Command
board.
5. The XJ6 and XJ7 connectors may still have a residual voltage of 35V DC from the C2 capacitor
located behind the Generator Supply Command board. Using a voltmeter (rating 1000 V DC), check
that there is no residual voltage (35V DC) between the XJ6 and XJ7 connectors.
S2
S1
RC2
RC1
XJ6
XJ7
C1P
M1P