Chapter 9
Page no. 1180
JC-DR-A-205.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A205 - Lift Motor
5
PREREQUISITES
•
Raise the column as required for access to the lift motor (1).
•
Switch off the system power.
•
Remove the front and the column covers (refer to
A044 - Remove/Reinstall Gantry Covers
).
•
Loosen and remove the Gantry curtain (refer to
6
PROCEDURE
6-1
Disassembly of the Lift Motor
1. Release the bolts securing the lower X plate
(2) (10 mm open-ended wrench)
2. Remove the curtain roller (3).
1
2
3