Chapter 9
Page no. 978
JC-DIAG-A-001.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card DIAG A001 - AWS Sun-VTS Diagnostics
3
PROCESSOR TESTS
3-1
Short Test
•
Type of test:
SUN VTS
For both V1/V2 and V3 Control Stations, from the command window, enter:
/opt/SUNWvts/bin/fputest –v –n
3-2
Long Test
•
Type of test:
SUN VTS
For both V1/V2 and V3 Control Stations, from the command window, enter:
/opt/SUNWvts/bin/fputest –v –f
4
HARD DRIVE TESTS
V1/V2 Control Stations have two hard disks with device names
c0t0d0
and
c0t1d0
. V3 Control Stations
have one large hard disk with device name
c1d0
.
4-1
Short Test
•
Type of test:
SUN VTS
For V1/V2 Control Stations to perform a short test on hard disk with device name
c0t0d0
, from a
command window, enter:
/opt/SUNWvts/bin/disktest –v -n -o dev=c0t0d0
For V1/V2 Control Stations to perform a short test on hard disk with device name
c0t1d0
, from a
command window, enter:
/opt/SUNWvts/bin/disktest –v -n -o dev=c0t1d0
For V3 Control Stations to perform a short test on hard disk with device name
c1d0
, from a command
window, enter:
/opt/SUNWvts/bin/disktest –v -n -o dev=c1d0
4-2
Long Test
•
Type of test:
SUN VTS
For V1/V2 Control Stations to perform a long test on hard disk with device name
c0t0d0
, from a com-
mand window, enter:
/opt/SUNWvts/bin/disktest –v -f -o dev=c0t0d0
For V1/V2 Control Stations to perform a long test on hard disk with device name
c0t1d0
, from a com-
mand window, enter:
/opt/SUNWvts/bin/disktest –v -f -o dev=c0t1d0
For V3 Control Stations to perform a long test on hard disk with device name
c1d0
, from a command
window, enter:
/opt/SUNWvts/bin/disktest –v -f -o dev=c1d0