GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A020 - Declare and Set Printers (Filming Devices)
Page no. 587
Chapter 7
JC-ELE-A-020.fm
Refer to
for information on Konica 793/873, Agfa 4500M/5500/5503/
AXYS, Kodak Dryview 6800/8900/8600/8610,, Carestream 5850, Carestream Dryview 6850, Fuji
FMDPL and Fuji Drypix 4000/5000/7000 printers.
Note:
Before using the values quoted below to configure a printer, contact the printer manufacturer FE
to check the software version installed on the printer.
6-3
Set Print Annotation Level
Ask the customer which print annotation model they want. Depending on the customer's requirement,
set the Printing Annotations selection to the one selected, using the Annotations window in Medical Pref-
erences. Refer to the Browser chapter in the
Senographe
DS
Operator Manual
for access to this win-
dow.