Chapter 9
Page no. 896
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
132/028
INHIBIT
Tilt overcurrent
A motion over current is detected.
Refer to
JC DIAG A1305 TILT CONTROL FAILURE
and to
to fix the problem.
A1305
D/R A292
132/029
INHIBIT
CPU HEARTBEAT
NOT RECEIVED BY
TILT NODE
A heartbeat failure occurred. The Gantry CPU did
not send periodic "heartbeat" CAN message.
1. Change the Gantry CPU Board (see
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration
).
3. Perform a Collimation Check (see
D/R A190
CAL A029
ELE A028
132/024
DEBUG
ONLY
HEARTBEAT
DISABLED ON TILT
NODE
The Gantry CPU tried to disable the "heartbeat"
check of the tilt node.
1. Change the Gantry CPU Board (see
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration
).
3. Perform a Collimation Check (see
D/R A190
CAL A029
ELE A028
132/030
DEBUG
ONLY
TILT INVALID
RX_OPTICAL_FOR
K1_ERROR
Ignore this error. If the problem persists, contact your
OLC.
132/031
DEBUG
ONLY
TILT INVALID
RX_OPTICAL_FOR
K2_ERROR
Ignore this error. If the problem persists, contact your
OLC.
132/032
DEBUG
ONLY
TILT INVALID
LIFT_DOWN_PEDA
L_ERROR
Ignore this error. If the problem persists, contact your
OLC.
132/033
DEBUG
ONLY
TILT INVALID
LIFT_UP_PEDAL_E
RROR
Ignore this error. If the problem persists, contact your
OLC.
132/034
DEBUG
ONLY
TILT INVALID
COMPRESSION_PE
DAL_ERROR
Ignore this error. If the problem persists, contact your
OLC.
132/035
DEBUG
ONLY
TILT INVALID
DECOMPRESSION
_PEDAL_ERROR
Ignore this error. If the problem persists, contact your
OLC.
132/036
DEBUG
ONLY
TILT INVALID
STOP_MOTION_MC
U_ACTIVE ERROR
Ignore this error. If the problem persists, contact your
OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card