2-7-5.
Supply Command Board 200PL2 Test Points
Test Point
Label (in French)
Function (in English)
PT1
Inib Decomptage
Inhibit counting B11
PT2
Contact K2
K2 Contact
PT3
Alim Puissance OK
Power Supply OK
PT4
Pres. Alim
Supply Present
PT5
250 V < E < 700 V
250 V < E < 700 V
PT6
Cde Contacteur
K1 Power Switch Command
PT7
MST
Power ON
PT8
Teta_Sup_Tmax
Overtemperature
PT9
RAZ 3 s
Reset Counter B11
PT10
Demande MST
Power ON Request
PT11
T_Sup_3S
Delay > 3 s
PT12
7.63 Hz
7.63 Hz Clock
PT13
Echauffement
Clock Counter B8
PT14
Auto MST
Automatic Power ON Signal
PT15
B13 Output
B13 Counter Outpuy
PT16
15.6 kHz Clock
15.6 kHz Clock
PT17
Contacteur
K1 Switch Command
PT18
+VLC (+12 V)
+12 V
PT19
Cde Precharge
Precharge DC Bus Command
PT20
Threshold pres acceleration
+ 174 mV DC Voltage
PT21
Threshold lanceur actif
+20 mV DC Voltage
P250V
+250 V
+250 V present
P27V
+27 V
+27 V present
0V/250V
0 V (250 V)
Ground for 250 V supply
0V/27V
0 V (27 V)
Ground for 27 V supply