ERR SUB A003 - Gantry Error Messages
141/006
INHIBIT
Coll out of FOV
(front)
Invalid parameter for axis[i] (for i=1 - 5) = collimator
control board received a command for axis[i] that has
an invalid parameter. Either a wrong FOV value or a
wrong filter type.
Possible causes of error:
- Collimator has received bad data on its node
- Communication conflict/mismatch between
system and collimator
Actions to correct problem:
1. Check the cable connection (W409) between the
collimator and system, then restart system from
the X-ray Console.
2. If the problem persists, change the CPU board
(re-calibrate the collimator afterwards).
3. If the problem persists, contact your OLC.
D/R A190
CAL A029
141/007
INHIBIT
Coll out of FOV (left)
141/008
INHIBIT
Coll out of FOV(rear)
141/009
INHIBIT
Coll out of FOV
(right)
141/010
INHIBIT
Coll filter type not
exist
141/011
INHIBIT
Coll beyond phy limit
(front)
axis[i](for i=1 - 4) beyond physical limit = the FOV is
beyond the blade capability.
1. Re-perform the collimator calibration. If the prob-
lem persists, identify the type of collimator cur-
rently in your system.
2. Replace the Collimator Set (refer to
3. Perform Collimator Format Calibration.
D/R A280
CAL A029
141/012
INHIBIT
Coll beyond phy limit
(left)
141/013
INHIBIT
Coll beyond phy limit
(rear)
141/014
INHIBIT
Coll beyond phy limit
(right)