Chapter 9
Page no. 1114
JC-DR-A-150.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A150 - System Fuses
that there is no residual voltage (250V DC) between the C1P and M1P connectors.
WARNING
Residual voltage stored in the C1 and C2 capacitors (located behind the Generator Com-
mand Supply board can be dangerous. Ensure that the capacitors are fully discharged
before you handle the fuses on the Generator Supply Command board.
C2
C1
XJ6
XJ7
C1P
M1P