ERR SUB A003 - Gantry Error Messages
101/001
DEBUG
ONLY
LIFT SOFTWARE
ERROR
On every command sent by the PTB, the IPMS
checks whether the command identifier is part of the
list of commands to be taken into account by IPMS. If
the command is not recognized, the command is
ignored, and “CommandIdError” is returned.
Should not occur in the field.
It could be a disconnect between the CPU soft
release and the Node soft release. Contact your
OLC if the problem persists.
101/002
DEBUG
ONLY
LIFT SOFTWARE
ERROR
On every command sent by the PTB, the IPMS
checks whether all parameters values are compati-
ble with their corresponding range. If not, the com-
mand is be ignored, and “ParameterError” is
returned to the PTB.
Should not occur in the field.
It could be a disconnect between the CPU soft
release and the Node soft release. Contact OLC if
the problem persists.
101/003
WARNING
LIFT CALIBRATION
ERROR
If the calibration is not achieved, for each CAN mes-
sage which is not a calibration command, the IPMS
ignores the command and sends the “CalibrationNot-
Achieved” error.
Use
Job Card CAL A031 - Lift Calibration
to perform a Lift calibration.
CAL A031
101/004
WARNING
NO VALID LINE
FOR LIFT
There is no valid line. The Gantry CPU requests a
motion but node does not see the valid line. The
valid line is an RT line, that is activated by the CPLD
of the node containing the UIF during the time when
the UIF is p a deceleration time. This does
not concern decompression and tilt.
This also occurs when the valid line falls before the
end of the movement.
Valid line problem. check the different motions, using
all keypads to determinate the defective node then
refer to
JC DIAG A1015 LIFT CONTROL FAILURE
A1015
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card